Placing spinal cord stimulator (SCS) leads in the epidural space carries a small risk of spinal cord injury, which may be higher if the spinal canal is narrowed. A preprocedure MRI can reveal stenosis or other problems at the planned lead entry or placement sites. A 2025 study in Neuromodulation examined how often such MRIs change the plan.
This summary is based on the published abstract. Read the abstract on PubMed
What the authors did
The authors reviewed 340 patients at one center who were considering a percutaneous SCS trial and had a preprocedure MRI. They assessed the location and severity of stenosis and analyzed whether MRI findings changed trial plans and whether patient characteristics predicted stenosis.
What they found
- MRI influenced SCS treatment in 7% of patients (25 of 340).
- Of these, 60% had the trial technique changed, and 40% did not go on to a trial because of an MRI finding.
- MRI was about 4.6 times more likely to change the plan for cervical trials than for thoracic or lumbar trials.
- Age was associated with moderate or severe stenosis, but patient characteristics were poor at predicting stenosis overall.
The authors recommend considering preprocedure MRI for all patients, examining both the lead entry and placement zones.
Why this matters
Because stenosis could not be reliably predicted from patient characteristics, imaging is the best way to find it. Identifying a narrow canal before placing leads can prevent serious neurological injury.
Limitations
This was a retrospective study from one center, and the study could not measure how many injuries were actually prevented.
Safety takeaways: how to apply it
For pain physicians and procedure teams
- Obtain and personally review a recent MRI of the lead entry and placement zones before SCS trials, especially cervical trials.
- Modify the entry level or approach, or reconsider the trial, when significant stenosis is present.
For hospitals, clinics and safety leaders
- Include MRI review in the pre-trial neuromodulation checklist.
For patients and families
- Ask whether you need an updated MRI before a stimulator trial.
Reference
- Provenzano DA, Vaidya EA, Kilgore JS. Preoperative Magnetic Resonance Imaging Modifies Percutaneous Spinal Cord Stimulator Trial Progression and Planning. Neuromodulation. 2025 Feb;28(2):306-314. doi:10.1016/j.neurom.2024.10.005. PMID: 39641705. PubMed: https://pubmed.ncbi.nlm.nih.gov/39641705/