CT Fluoroscopy Used Far Less Radiation Than Spiral CT for Lumbar Epidural Injections

CT Fluoroscopy Used Far Less Radiation Than Spiral CT for Lumbar Epidural Injections

By Ivan Urits, MD – Board Certified in Anesthesiology and Pain Medicine

Published March 24, 2026 | Category: Pain Medicine

Some lumbar epidural steroid injections are guided by CT, especially in complex anatomy. CT can be used in different ways. "Spiral" CT takes repeated full scans, while CT fluoroscopy gives quick, near-live images of a small area. A study in Skeletal Radiology compared how much radiation patients received with each method.

This summary is based on the published abstract. Read the abstract on PubMed

What the authors did

The authors reviewed 767 consecutive lumbar epidural injections done at their center from 2015 to 2023: 240 with CT fluoroscopy and 527 with spiral CT. They compared patient radiation, procedure time, pain relief and complications, including in cases done by attending physicians alone and by trainees.

What they found

  • Patient radiation was about 85% lower with CT fluoroscopy: 37.2 vs 251.1 mGy·cm.
  • Procedures were faster: 20.1 vs 29.6 minutes.
  • Immediate pain relief was similar between groups.
  • There were no accidental intrathecal punctures with CT fluoroscopy, compared with four in the spiral CT group.
  • The radiation and time savings held for both attending-only and trainee-performed procedures.

Why this matters

Many patients receive several spine injections over the years, and radiation exposure adds up. When CT guidance is needed, how it is used matters. Switching technique gave large reductions in radiation and time without losing effectiveness.

Limitations

This was a retrospective study from one practice, and patients were not randomly assigned. The comparison does not apply directly to standard C-arm fluoroscopy, which is how most epidural injections are guided.

Safety takeaways: how to apply it

For pain physicians and procedure teams

  • When CT guidance is needed, consider CT fluoroscopy and low-dose protocols rather than repeated spiral scans.
  • Limit the scanned area and the number of images to what is needed.
  • Track dose for every procedure, including those done by trainees.

For hospitals, clinics and safety leaders

  • Review CT-guided spine injection protocols with radiology and medical physics to minimize dose.

For patients and families

  • Ask what kind of imaging will guide your injection and how radiation will be kept low, especially if you have had many prior scans.

Reference

  1. Shiang T, Wieschhoff GG, Wu JS, Mandell JC. Comparison of patient radiation exposure and procedure time between CT-fluoroscopy and spiral CT for lumbar epidural steroid injections. Skeletal Radiol. 2025 Jul;54(7):1491-1501. doi:10.1007/s00256-024-04858-2. PMID: 39710723. PubMed: https://pubmed.ncbi.nlm.nih.gov/39710723/

This article is educational information, not legal or medical advice.

Previous
Previous

Safety Practices for Disc Procedures, Kyphoplasty and Neurostimulator Implants

Next
Next

Fibrin Glue Sealed Accidental Dural Punctures During Stimulator Lead Placement