Ultrasound Probe Covers Can Tear Without Anyone Noticing

Ultrasound Probe Covers Can Tear Without Anyone Noticing

By Ivan Urits, MD – Board Certified in Anesthesiology and Pain Medicine

Published December 2, 2025 | Category: Pain Medicine

Sterile covers protect ultrasound probes and patients during needle procedures, including many pain procedures. Some cleaning protocols assume the cover stays intact, allowing lower-level disinfection of the probe afterward. A 2025 study in the Australasian Journal of Ultrasound in Medicine tested whether covers are actually intact after use.

This summary is based on the published abstract. Read the abstract on PubMed

What the authors did

In a single emergency department, the authors collected two types of transducer covers, made of polyethylene or polyurethane, after ultrasound-guided needle procedures. They tested each cover for holes by filling it with a liter of water and checking for leaks, and recorded details about the patient, procedure and operator.

What they found

  • Polyethylene covers were perforated after 15.1% of procedures, more than six times the rate of polyurethane covers (2.4%).
  • Operators were largely unaware that the covers had been perforated.

The authors conclude that relying solely on the cover for infection control appears unwise.

Why this matters

If a cover tears during a procedure, the probe may become contaminated, and if it is not properly disinfected, it could spread infection to the next patient. This study supports cleaning protocols that do not assume the cover stayed intact.

Limitations

This was a single-center study in an emergency department, using covers from one manufacturer. It did not determine when the holes occurred or whether probes were actually contaminated.

Safety takeaways: how to apply it

For pain physicians and procedure teams

  • Do not assume a probe cover remained intact; follow full disinfection protocols after needle procedures.
  • Choose more durable cover materials when possible.
  • Inspect covers after use, and report tears.

For hospitals, clinics and safety leaders

  • Review probe reprocessing policies, and consider cover material when purchasing supplies.

For patients and families

  • It is reasonable to ask how ultrasound equipment is cleaned between patients.

Reference

  1. Rippey J, Morris F, Moseley C, Sowerby J, Chiverton L, Mason J. Post-Procedure Perforation Rates in Ultrasound Transducer Covers Used for Ultrasound-Guided Percutaneous Procedures in Emergency Medicine: A Comparison Between Polyurethane Covers and Polyethylene Covers. Australas J Ultrasound Med. 2025 Nov;28(4):e70025. doi:10.1002/ajum.70025. PMID: 41017784. PubMed: https://pubmed.ncbi.nlm.nih.gov/41017784/

This article is educational information, not legal or medical advice.