Broader MRI Labeling Helped More Stimulator Patients Get Scans Within Safety Conditions

Broader MRI Labeling Helped More Stimulator Patients Get Scans Within Safety Conditions

By Ivan Urits, MD – Board Certified in Anesthesiology and Pain Medicine

Published February 17, 2026 | Category: Pain Medicine

Many people with spinal cord stimulators (SCS) will need an MRI at some point. Each device is approved for MRI only under specific conditions, such as the scanner strength, the body area scanned and where the leads sit. When a needed scan falls outside those conditions, it can only be done "off-label," if at all. A study in Pain Practice asked whether an expanded MRI approval for one device let more scans be done within the manufacturer's recommendations.

This summary is based on the published abstract. Read the abstract on PubMed

What the authors did

The authors reviewed all MRI scans in patients with one stimulator system (Abbott Proclaim) at a German center. They used the technical data saved with each scan and the patients' treatment records. Because the hospital changed its MRI procedures in 2019, the periods before and after 2019 were analyzed separately.

What they found

  • 62 MRI scans were performed in patients with the device.
  • The full implanted system was MRI-approved in just over 50% of cases, under both the old and new approval.
  • The expanded lead placement options did not significantly increase the number of MRI-conditional systems.
  • However, the new approval allowed significantly more scans within the recommended settings for a higher energy level (specific absorption rate) in the later period.
  • No device-related adverse events were seen.
  • Cervical leads remained a problem, because they require longer leads and lower impedance can prevent the device from entering MRI mode.

Why this matters

MRI with an implanted stimulator can be done safely, but only when the device, leads and scan settings match the approved conditions. Choosing an MRI-friendly system and planning lead placement at the time of implant can save patients from being denied a needed scan later.

Limitations

This was a small, retrospective review of a single device at one center. Findings may not apply to other manufacturers' systems, and approvals continue to change.

Safety takeaways: how to apply it

For pain physicians and procedure teams

  • Consider future MRI needs when choosing a stimulator system and planning lead placement, especially for cervical systems.
  • Give every patient documentation of their exact device and leads and the MRI conditions that apply.
  • Confirm the device is placed in MRI mode before any scan.

For hospitals, clinics and safety leaders

  • Create a shared protocol between radiology and the pain team for screening and scanning patients with implanted stimulators.

For patients and families

  • Carry your device ID card, and tell the MRI team about your stimulator well before the scan date.

Reference

  1. Reining M, Winkler D, Kirchhof K, Boettcher J, Kretzschmar M. Effects of an extended MRI approval of an implantable spinal cord stimulation device on compliance with manufacturer's recommendations. Pain Pract. 2025 Apr;25(4):e70022. doi:10.1111/papr.70022. PMID: 40159636. PubMed: https://pubmed.ncbi.nlm.nih.gov/40159636/

This article is educational information, not legal or medical advice.