Hardware Complications of Spinal Cord Stimulators: What Patients and Physicians Should Know
By Ivan Urits, MD – Board Certified in Anesthesiology and Pain Medicine
Published March 25, 2025 | Category: Pain Medicine
Spinal cord stimulators (SCS) relieve nerve pain using thin wires (leads) placed in the epidural space and a battery-powered generator implanted under the skin. They are minimally invasive and reversible, but the hardware can fail. A 2024 review in Current Pain and Headache Reports summarized device-related complications of cylindrical-lead systems.
This summary is based on the published abstract. Read the abstract on PubMed
What the authors did
The authors reviewed recent studies on device-related SCS complications, covering their causes, symptoms, diagnosis, clinical impact and treatment options.
What they found
- Device-related complications are more common than biological ones such as infection or bleeding.
- The main device problems include:
- Lead migration: the lead moves from its intended position, and pain relief may be lost.
- Lead fracture or disconnection: often causing sudden loss of stimulation or abnormal sensations.
- Generator failure or loss of charge.
- Generator flipping in its pocket, which can prevent charging or programming.
- Hardware-related pain at the generator or incision site.
- Paresthesia intolerance: the tingling sensation from stimulation is uncomfortable.
The authors stress that considering these complications before trials and implants is an important part of patient management and device selection.
Why this matters
Device complications often lead to revision surgery and loss of pain relief. Knowing their patterns helps physicians choose devices and techniques, explain realistic risks and recognize problems early.
Limitations
This is a narrative review rather than a systematic analysis, so it does not provide pooled complication rates.
Safety takeaways: how to apply it
For pain physicians and procedure teams
- Use anchoring techniques and strain-relief loops to reduce lead migration and fracture.
- Place the generator in a location that limits flipping and discomfort.
- Discuss the possibility of revision surgery during consent, and evaluate sudden loss of stimulation with imaging and device checks.
For hospitals, clinics and safety leaders
- Track revisions and explants by cause to identify device or technique problems.
For patients and families
- Report sudden changes in stimulation, new pain at the device site or charging problems to your implanting physician.
Reference
- Hasoon J, Vu PD, Mousa B, Markaryan AR, Sarwary ZB, Pinkhasova D, et al. Device-Related Complications Associated with Cylindrical Lead Spinal Cord Stimulator Implants: A Comprehensive Review. Curr Pain Headache Rep. 2024 Sep;28(9):941-947. doi:10.1007/s11916-024-01280-0. PMID: 38850491. PubMed: https://pubmed.ncbi.nlm.nih.gov/38850491/
This article is educational information, not legal or medical advice.